Connecting the EASY-Spray Column Plumbing
Using the nanoViper Fitting
22
EASY-Spray Series Ion Source User Guide
Thermo Scientific
ii. In the EASY-Spray dialog box, press the temperature box, enter a temperature
value in the range 30–60 °C, press
Accept
, and then press
Set
and
Close
.
–or–
• (RSLCnano system) Use the temperature control knob on the front of the ion source.
lists the display readings that can appear on the actual temperature display.
Using the nanoViper Fitting
For some LC instruments, the plumbing connects to the EASY-Spray column through several
nanoViper fittings, such as the one shown in
. Although these fittings can withstand
ultra-high-performance LC (UHPLC) backpressures up to 1200 bar (17 400 psi), they are
fingertight fittings that require only very small torques to seal. Therefore, you must follow the
next procedure to avoid damage by overtightening.
Figure 12.
nanoViper fitting
IMPORTANT
If you later manually turn the temperature control knob, you
must repeat this procedure to reset the touch-screen value.
Table 5.
Readings for the actual temperature display
Display
Description
--
EASY-Spray column is not connected.
xx
.
Actual temperature while heating up
xx
Actual temperature, stabilized
b.L.
Disconnect and reconnect power to the ion source to exit boot
loader mode (b.L.).
No display
Replacing the Temperature Controller PCB
Tip
Refer to the packing for your nanoViper fittings to determine compatibility with
either the 1000 bar capillaries (beige tubing) or the 1200 bar capillaries (blue tubing).
PEEK sealing surface
Removable knurled tightening tool
Screw