2
Installing the EASY-Spray Series Ion Source
Adjusting the Emitter Tip Position
12
EASY-Spray Series Ion Source User Guide
Thermo Scientific
Adjusting the Emitter Tip Position
Before installing the EASY-Spray column, use the emitter positioning tool (
3 mm hex key to adjust the position of the column holder.
shows the emitter positioning tool touching the MS ion transfer tube (0 mm gap).
The length of the tool’s tip ensures the 3 mm gap for the EASY-Spray column’s emitter.
Figure 7.
Emitter positioning tool
Figure 8.
Gap positions in front of the MS ion transfer tube (side view)
0 mm gap between the ion transfer tube
and the emitter positioning tool
3 mm gap between the ion transfer tube
and an actual emitter
MS spray cone
MS ion transfer tube
EASY-Spray column
(shown without its cable)
Emitter positioning tool