1
–
Introduction
Thermo Scientific
Product Manual for IonPac AS31 Columns
Page 9 of 45
155007-01
For Research Use Only. Not for use in diagnostic procedures.
Table 1
Dionex IonPac AS31/Dionex IonPac AG31 Operating Parameters
Column(s)
Typical Back Pressure
psi (MPa
a
), 15°C
b
Standard
Flow Rate
mL/min
Maximum
Flow Rate
mL/min
c
Dionex IonPac AS31
2 mm Analytical column
~ 3000 (20.68)
0.30
0.40
Dionex IonPac AG31
2 mm Guard column
~ 200 (1.38)
0.30
0.40
Dionex IonPac AG31 and AS31
2 mm columns
~ 3200 (22.06)
0.30
0.40
Note: 1MPa = 145.04 psi
b
Total backpressure at standard flow rates
c
In all cases, flow rate should not result in pressures over 5,000 psi
For assistance, visit Unity Lab Services online at www.unitylabservices.com.
From the U.S., call the Customer Care Center for Dionex Products at 1-800-346-6390.
Outside the U.S., call the nearest Thermo Fisher Scientific office.
NOTE
!