Alarm Action
The Alarm Action dropdown list allows the user to make changes to alarms as a group,
either system-wide or by application.
–
Do Nothing – No action is taken when sent to the gauge.
–
Clear all alarms for all applications – Currently this option is non-functional and
no action is taken upon its selection. This option will be utilized in future versions
of the EZ Cal II software.
–
Disable all alarms – Any process alarm or mode/fault alarm that is in an enabled
state is changed to the inhibit state. This option is available for all applications or
individual applications.
–
Enable all alarms – Any process alarm or mode/fault alarm that is in the inhibit
state is changed to an enabled state. This option is available for all applications or
individual applications.
–
Erase all process alarm assignments – All configurations associated with setting up
a process alarm are set to the default values. These values include Measurement
ID, Action, Action Delay, Set Point, Clear Point, and the Enable/Disable setting.
This option is available for all applications or individual applications.
–
Erase all mode/fault alarm assignments – All configurations associated with setting
up a mode/fault alarm are reset to the default value. Enable/Disable setting,
Action. This option is available for all applications or individual applications.
–
Erase all alarm assignments, all mode/fault/process alarms for all applications –
This option allows the user to perform the Erase all process alarm assignments
action and the Erase all mode/fault alarm assignments action simultaneously. This
option is not available for individual applications.