Upgrading CPU firmware from
Rev <= 3.500 to version => 3.6
This
procedure
does not
apply
Note the current Primary Measurement Type and unit
displayed on the device
Reset the gauge Primary Measurement Type to
“Density” and units to “g/cc”
Upload CFG from Gauge using EZ CAL II and save the
file to your laptop
Perform the firmware upgrade
Download File to Gauge using EZ CAL II.
Use previously saved configuration file on your laptop
Manually re-enter the process variable units that you
noted earlier
End
Start
No
Yes
Upload CFG from Gauge using EZ CAL II and save the
file to your laptop for future reference
x
i
x
DensityPRO User Manual
Thermo Scientific
DensityPRO Firmware Upgrade Process
Note: When upgrading from CPU firmware versions of 3.500 or older to versions
3.600 or newer
,
there is the potential for a mismatch of the measurement units
to occur. In order to prevent this
from
happening
,
the following procedure must
be followed.