AutoMITTER SIB Option
ATEX or Brazilian Certified Systems
6-10 AutoPILOT PRO Flow Computer User Guide
Thermo Fisher Scientific
Read the following warnings before making any connections.
Warning
Do not install the system in any hazardous area, except in those
for which the system has been approved.
Warning
Do not make any connections unless power has been removed
and the area is known to be non-hazardous. Once connections are made,
do not disconnect unless power has been removed and the area is known to
be non-hazardous.
Warning
Substitution of components may impair the intrinsic safety of the
product.
1.
Open the enclosure door. Loosen the thumbscrews on the mounting
bracket, and lift the main board and mounting bracket up to expose the
chassis bracket.
2.
Refer to Figure 6–6 for the following.
a.
Mount the SIB onto the standoffs on the SIB mounting bracket.
b.
Mount the SIB assembly onto flow computer chassis bracket.
Tighten the fasteners to secure the SIB assembly.
Figure 6–8. Installing the SIB onto the bracket
3.
TB14 on the main board is the SIB connector. Using the SIB cable,
connect TB14 to TB5 on the SIB as shown in
Installation