Installing the Basic System
Mounting the Enclosure onto the Transducer
3-4 AutoPILOT PRO Flow Computer User Guide
Thermo Fisher Scientific
6. Secure the transducer to the enclosure using the sealing washers and
bolts.
Note
Before tightening the bolts, connect the ground wire to the bolt
closest to the ground lug in the enclosure.
Figure 3–3. Direct mounting steps 5–6
7. Connect the ground wire to the ground lug in the enclosure.
8. Loosen the thumbscrews on the mounting bracket, and lift the main
board and mounting bracket up. Route the transducer cable through
the opening in the mounting bracket that provides access to J25 on the
back of the main board. If necessary, remove the bracket over the
opening, route the cable, and then replace the bracket. Connect the
transducer cable to J25 on the main board. (J25 is shown in
Figure C–2.
)