8. Troubleshooting
Problem
Cause
Solution
Power on, but no display
a)
Loose connections
b)
Incorrect output voltage of
the power adaptor
c)
Cables not in correct polarity
(+ and – position)
a)
Ensure cables make good
contact
b)
Use an power adaptor with
specified output voltage
c)
Re-wire loop cables with correct
polarity
Unstable conductivity
reading
a)
Air bubbles in the probe
b)
Dirty probe
c)
Probe not deep enough in
sample
a)
Tap probe to remove air bubbles
b)
Clean probe and recalibrate
c)
Ensure that the sample entirely
covers the sensor of the probe
Oscillating temperature
readings
a)
Electrical noise interference
a)
Ensure shield wire is properly
connected to pin 7
Slow response
a) Dirty / Oily probe
a) Clean probe
Blinking ATC
a)
No temperature probe
connection during ATC mode
a)
Ensure temperature sensing
cable makes good contact
Blinking electrode
annunciator
a)
Error in calibration
b)
Conductivity value of the
calibration standard is not
within acceptable range
a)
Ensure calibration standard
solution is not contaminated.
Ensure electrode is clean
b)
Ensure calibration standard is
within 10 to 100% of full-scale of
the selected measuring range
Or (Conductivity)
a)
Conductivity reading is
above the selected
measuring range
a)
Ensure correct conductivity
measurement range is selected
Or/Ur (Temperature)
a)
Temperature probe is not
connected when ATC
enabled
b)
Temperature value exceeds
above/ below the specified
range
a)
Ensure electrode makes good
contact with monitor
b)
Ensure temperature calibration is
done properly. Ensure process
temperature is within the
specified range