Instruction Manual
Alpha-
COND 550
26
Calibration Default Settings (CAL dEF)
Conductivity
Calibration factor
Reset to 1.000
Factory Default Settings (FCt dEF)
Conductivity
Calibration factor
Reset to 1.000
Cell constant
1.00
Cell type
2-cell
Measuring range
0 to 2000 µS/cm
Line resistance
0.000
Ω
Temperature
Unit of measurement
o
C
Temperature compensation
ATC On
Temperature coefficient
2.1% (Both process & calibration temp. coefficients)
Temperature coefficient type
Linear
MTC (ATC Off)
25
o
C (Both process & calibration temperature)