Instruction Manual
Alpha-
COND 550
25
6. Technical Specifications
General Specification
Conductivity
Measuring Range:
Resolution:
0.00 - 20.00 µS/cm
0.01 µS/cm
0.0 - 200.0 µS/cm
0.1 µS/cm
0- 2000 µS/cm
1 µS/cm
0.00 - 20.00 mS/cm
0.01 mS/cm
0.0 - 200.0 mS/cm
0.1 mS/cm
Relative accuracy
± 1% of full scale reading
System accuracy
The effective measuring range and the total system
accuracy depend on the sensor used with the monitor.
Please refer respective sensor documentation.
Temperature
Measuring range
-10.0 to +125.0
o
C
+14.0 to +257.0
o
F
Resolution 0.1
o
C
0.1
o
F
Relative accuracy
±0.5
o
C
±1.0
o
F
Sensor
2 wire / 3 wire - Pt100 or Pt1000 (Jumper Selectable)
Compensation
Automatic or Manual (reference at 25
o
C)
Calibration
Conductivity
Number of calibration points
1 point
Calibration window
±40% of default reading
Calibration standard solution window 10% to 100% of the selected measuring range
Temperature
Offset Adjustment
± 10 ˚C/ ± 18 ˚F
Display
LCD
7 segments display with symbols for status information
Electrical Data and Connections
Conductivity Input
Screw Terminal (3.5mm pitch)
Connection terminal
8-pin & 2-pin terminal blocks, DC power connector
(1.2mm diameter)
EMC Specifications
Emission
EN 61 36
Susceptibility
EN 61326
Other
Power Input
Uni9 V DC Adaptor and 24V DC Supply
Dimensions (W x H x D)
96mm x 96mm x 66 mm
Weight (Estimated)
210g
Ambient Temp. operating range
0 to 40
o
C
Maximum Relative Humidity
80% up to 31
o
C decreasing linearly to 50% at 40
o
C