Instruction Manual
Alpha-
COND 550
20
If ATC enabled (ATC On):
4
Selecting temperature offset:
The upper display shows the last
configured temperature offset value (if any), otherwise the default is
zero. The lower display shows currently measured temperature
reading (including last configured offset value). LCD shows ‘
ATC
’
annunciator in lower-right corner.
Place a thermometer, which is known to be accurate, in your
sample or process liquid. Make sure your temperature probe is
placed in the same liquid. Compare the stabilized temperature
reading displayed on the monitor with the thermometer. If there is a
difference between the two readings (offset), you can adjust the
reading of the monitor.
Press
▲
or
▼
key
to adjust the lower
display to the correct temperature value.
5
5
As the lower display value changes, the monitor adjust the upper
display reading automatically to suit the new offset value. Up to ± 10
ºC/ ± 18 ºF offset is allowed.
Press ENT key
to confirm the value. The monitor reverts to
SET
ºCF
screen.
If ATC disabled (ATC OFF):
6
Setting process temperature
: The lower display shows ‘
P.ºC
’ and
the upper display shows the last configured process temperature.
Press
▲
or
▼
key
to adjust the upper display to desired process
temperature. Allowable range: –10.0 to 125.0°C / 14.0 to 257.0°F.
Press ENT key
to confirm the process temperature.
7
Setting calibration temperature
: The lower display shows ‘
C.ºC
’
and the upper display shows the last configured calibration
temperature.
Press
▲
or
▼
key
to adjust the upper display to
desired calibration temperature. Allowable range: –10.0 to 125.0°C /
14.0 to 257.0°F.
Press ENT key
to confirm the calibration temperature. The monitor
reverts to
SET ºCF
screen.
Press
▲
or
▼
key
to access other setup screens or
press
▲
and
▼
key
simultaneously
(escape) to return to measurement mode.
To exit from any intermediate steps,
press
▲
and
▼
keys simultaneously
(escape). The monitor returns to the first screen of temperature settings
SET
ºCF.