Benchtop pH/ISE Meter Instruction Manual
Troubleshooting
76
Problem
Probable Cause
Remedy
Trouble
Some electrodes
Perform a manual calibration
Calibrating in
may operate outside
Autocalibration
the limits of ± 0.5
Buffers used may be
pH units.
out of specification.
Repeat calibration
using fresh buffers or from a
different lot.
Meter display
Meter failure
For Models 420Aplus,
Locks
710Aplus, reset meters as
described in set-up menu 2-4.
For Models 520Aplus,
525Aplus, 720Aplus
920Aplus, unplug meter,
hold 0 key down and plug
meter in. Press yes at the
prompt “RESET METER?”