![Thermo orion 410Aplus Instruction Manual Download Page 127](http://html1.mh-extra.com/html/thermo-orion/410aplus/410aplus_instruction-manual_1104441127.webp)
710Aplus
720Aplus
920Aplus
pH range
-2.000 to 19.999
-2.000 to 19.999
-2.000 to 19.999
resolution
0.001/0.01/0.1pH
0.001/0.01/0.1pH
0.001/0.01/0.1pH
relative accuracy
± 0.005
±0.002
±0.002
slope
80 to 120%
80 to 120%
80 to 120%
concentration
range
0.000 to 19900
0.000 to 19900
0.000 to 19900
resolution
± one least significant digit
± one least significant digit
± one least significant digit
relative accuracy
±0.05% of reading
±0.5% of reading
±0.5% of reading
temperature
-5.0 to 105.0 oC
-5.0 to 105.0 oC
-5.0 to 105.0 oC
resolution
0.1 oC
0.1 oC
0.1 oC
relative accuracy
±1.0 oC
±1.0 oC
±1.0 oC
millivolt range
-1600.0 to +1600.0
-1600.0 to +1600.0
-1600.0 to +1600.0
resolution
0.1 mV
0.1 mV
0.1 mV
relative accuracy
±0.2 mV or ±0.05% of readings
±0.2 mV or ±0.05% of reading
±0.2 mV or ±0.05% of reading
whichever is greater
whichever is greater
whichever is greater
relative millivolt
range
-1999.9 to +1999.9
-1999.9 to +1999.9
-1999.9 to+1999.9
resolution
0.1 mV
0.1 mV
0.1 mV
relative accuracy
±0.2 mV or ±0.05%
±0.2 mV or ±0.05%
±0.2 mV or ±0.05%
of reading whichever is greater
of readings whichever is greater
of reading whichever is greater
display
Custom LCD
Custom 2-color
Custom 2-color
vacuum fluorescent
vacuum fluorescent
with 17 character prompt line
with 17 character prompt line
inputs
two BNC, two pin tip
two BNC, two pin tip
two BNC, two pin tip
ATC, power, RS232
ATC, power, RS232, Karl Fischer
ATC, power, Karl Fischer
outputs
RS232
RS232, 2.5 mm
RS232, 2.5 mm
pin recorder,
pin recorder
power
9 VDC line converter for either 110 9 VDC line converter for either 110
9 VDC line converter for either 110
requirements
or 220 VAC. 50/60 Hz walloutlet.
or 220 VAC. 50/60 Hz wall outlet.
or 2209 VAC. 50/60 Hz waloutlet.
input
>10
12
ohms
>10
12
ohms
>10
12
ohms
impedance
instrument drift
< 50 microvolts/oC
< 50 microvolts/oC
< 50 microvolts/oC
input bias current
< ± 1 picoamp at 25oC and < ± 4 picoamps over full operating range
environmental
5 to 45 oC and 5 to 85% relative humidity, non-condensing
requirements
meter dimensions
8.08 x 3.26 x 1.90 inches
case
Dust and Splash Resistant, Chemical Resistant
SPECIFICATIONS
Specifications
126
Summary of Contents for 410Aplus
Page 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Page 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Page 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Page 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Page 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Page 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...