![Thermo orion 410Aplus Instruction Manual Download Page 113](http://html1.mh-extra.com/html/thermo-orion/410aplus/410aplus_instruction-manual_1104441113.webp)
Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions 520Aplus,525Aplus,720Aplus and 920Aplus
112
Transfer Data
To transfer data from the internal datalogger to a printer or computer press
yes in response to the TRANSFER DATA? prompt. While the data is being
transferred the prompt line will display DATA TRANSFER. On completion
the DATA TRANSFERRED prompt will be displayed followed by the next
option in the menu.
Data Clear
Transferring data from the datalogger does not clear the memory. Pressing
yes to the DATA CLEAR? prompt will clear the datalogger memory. DATA
CLEARING followed by DATA CLEARED will be displayed. The meter
then displays the next option in the menu, AUTOSHUTOFF.
NOTE: Prior to using the datalogger for the very
first time, it will be necessary to clear datalogger
memory with the DATA CLEAR option.
Autoshutoff Feature
The meters have a special feature that will help to prolong meter life called
autoshutoff. When autoshutoff has been activated the meter will enter
standby mode if no keys have been pressed for ten minutes. To start the
meter again, press any key.
Prompt
Keypress/Action
Press 2nd then Options.
1. OPTIONS MENU
Press no.
CHANGE ISO PT?
2. TRANSFER DATA?
Press no.
3. DATA CLEAR?
Press no
4. AUTOSHUTOFF ON
Press yes.
5. AUTOSHUTOFF OFF
The meter verifies your
selection then proceeds to the
next item in the options menu.
(The 520Aplus, 525Aplus, and
720Aplus return to
measure mode).
Summary of Contents for 410Aplus
Page 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Page 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Page 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Page 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Page 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Page 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...