Code Function
Description
2-3
ISOPOTENTIAL
Use to change the isopotential point
POINT
for a particular pH electrode. In pH mode
the default value is 7.00. The isopotential
point may not be changed in any other mode.
NOTE: Setup function 2-3 Isopotential point is
only accessible in pH mode. SETUP Functions
2-1 Slope and 2-2 Resolution are only accessible in pH
and concentration modes (710Aplus).
2-4
RESET
Allows all the setup options and calibration
data to be set to factory default values. This
is particularly useful during troubleshooting
or prior to calibrating with new electrode. To
RESET press ,ON will flash and the audio
signal will ring rapidly. Press yes.
3-1
TIMER
Sets the timer interval. The maximum
INTERVAL
interval that can be set is 23 hours, 59
minutes, and 59 seconds. The minimum
interval is five (5) seconds. When the
TIMER INTERVAL code, 3-1, is displayed
the current interval hours setting is displayed
in the main field (H 00). Press yes to accept,
or press to change, then press yes. Next the
current interval minutes: seconds will be
displayed (00:00). Press yes to accept current
setting or press to desired value, then press
yes. The default setting is 5 seconds.
3-2
TIME
Allows review of the time remaining before
the timer is set to go off.
3-3
SET REAL
Set the actual time of day. The meter TIME
uses a 24 hour clock. When the code 3-3
is displayed in the lower field, the current
time (hours:minutes) is displayed in the main
field. If correct press yes, otherwise change
as required and then press yes.
▼
Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions-420Aplus, 710Aplus
96
▼
▼
Summary of Contents for 410Aplus
Page 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Page 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Page 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Page 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Page 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...