Benchtop pH/ISE Meter Instruction Manual
Other Measurement Modes
49
B. Models 420Aplus, 710Aplus Millivolt Measurements
The Models 420Aplus and 710Aplus can be used to measure absolute or
relative millivolts. The millivolt modes are useful when performing
potentiometric titrations or preparing calibration curves. Detailed
instructions for any Thermo Orion electrode are given in the electrode
instruction manual. Titration instructions are included in the Thermo Orion
Redox Electrode (Model 9678 or 9778) Instruction Manual, or in standard
analytical chemistry texts.
Absolute Millivolts
Absolute millivolts are displayed with 0.1 mV resolution in the range of
-1600.0 to +1600.0 mV.
Access the absolute millivolt mode by pressing mode until the mV mode
indicator is displayed.
Relative Millivolts
Relative millivolts are displayed with 0.1 mV resolution over the range of
-1999.9 to +1999.9 mV. (Absolute millivolt range ± 1600.0).
1. Access relative millivolt mode by pressing mode until the REL mV
mode indicator is displayed.
2. Set the relative millivolt offset by pressing 2nd then cal.
CALIBRATE and the current absolute millivolts will be displayed.
Once the signal is stable, the meter displays 0.0. Use or to enter
the desired value, or leave the setting at 0.0. Press yes. The meter
automatically returns to MEASURE and all relative millivolt
measurements will be based on the offset.
▼
▼
Summary of Contents for 410Aplus
Page 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Page 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Page 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Page 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Page 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Page 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...