__________________ LCQ Deca XP Plus Hardware Manual __________________
5-1
Chapter 5
Service Maintenance
This chapter describes MS detector maintenance procedures. It is your
responsibility to maintain your system properly. However, the procedures in
this chapter are perhaps best performed by a Thermo Finnigan Customer
Support Engineer.
MS detector maintenance procedures are listed in Table 5-1.
*
Frequency depends on analytical conditions
Table 5-1.
Service maintenance procedures
MS Detector
Component
Procedure
Frequency
Procedure Location
Ion optics
Clean quadrupole,
octapole, and interoctapole
lens
As needed*
Trap entrance lens
Clean trap entrance lens
As needed*
Mass analyzer
Clean mass analyzer
Yearly (or as needed*)
Ion detection system
Clean ion detection system
(electron multiplier and
conversion dynode)
Whenever the top cover
plate of the vacuum
manifold is removed
Ion detection system
Replace electron multiplier
anode and cathode
If noise in spectrum is
excessive or proper
electron multiplier gain
can not be achieved
Fuses
Replace fuse
If fuse has blown
PCBs
Replace PCB
If PCB fails
Ion gauge
Replace ion gauge
If ion gauge fails
Turbomolecular
pump
Replace turbomolecular
pump
If turbomolecular pump
fails