![Thermo Finnigan LCQ Deca XP Plus Hardware Manual Download Page 100](http://html.mh-extra.com/html/thermo-finnigan/lcq-deca-xp-plus/lcq-deca-xp-plus_hardware-manual_1104316100.webp)
User Maintenance
API Source Maintenance ______________________________________________________
4-30
___________________ LCQ Deca XP Plus Hardware Manual ________________
5. Grasp the spray shield with both hands and carefully pull it and the API
stack free from the vacuum manifold. Place the API stack on a clean
surface with the spray shield down. Allow the API stack to cool to
ambient temperature before you disassemble the API stack.
The API stack is shown in Figure 4-8.
Go on to the next topic: Disassembling the API Stack.
Note. If you are unable to dislodge the spray shield from the vacuum
manifold, reattach the API flange to the spray shield and then pull the flange
away from the vacuum manifold.
Figure 4-8.
API stack
Spray Shield
Spray Cone
Retainer Clip
Extension Mount
Tube Lens
Tube Lens
And Skimmer
Mount