Service Maintenance
Replacing PCBs and Assemblies in the Tower _____________________________________
5-40
___________________ LCQ Deca XP Plus Hardware Manual ________________
d. Disconnect the cable that is connected to the fan cable.
e.
Disconnect the two cables to the Power Module from the two (upper)
connectors on the Power Module.
f.
Disconnect the cable that is connected to the reset button cable.
g. Loosen by hand (or with a Phillips screwdriver) the fastener that
holds the switching power supply assembly to the tower.
h. Remove the switching power supply assembly from the tower.
i.
Unpack the new switching power supply assembly
(P/N 97000-60151). Retain the packing materials so that you can
pack and ship the defective switching power supply assembly to the
Thermo Finnigan Repair Center in San Jose. Be sure to note the
apparent problem or symptoms on the enclosed forms.
j.
Install the new switching power supply assembly in the space
occupied by the old assembly.
k. Tighten by hand the fastener that holds the switching power supply
assembly to the tower.
l.
Reconnect the cable that is connected to the reset button cable.
m. Reconnect the two cables to the Power Module to the two (upper)
connectors on the Power Module.
n. Reconnect the cable that is connected to the fan cable.
o. Reconnect the cable to the Waveform Amplifier PCB to the connector
on the Waveform Amplifier PCB.
p. Reconnect the two cables to the Analyzer Auxiliary PCB to the two
connectors on the Analyzer Auxiliary PCB.
q. Reconnect the cable to the RF Voltage Amplifier PCB to the
connector on the RF Voltage Amplifier PCB.
r.
Go to step 7.
5. To replace the Turbomolecular Pump Controller, or the +48V, +24V
switching power supply, proceed as follows:
a.
Disconnect from the Turbomolecular Pump Controller the thick cable
to the turbomolecular pump.
b. Disconnect from the Turbomolecular Pump Controller the thin cable
that comes from the Power Module.
c.
Disconnect the four pin connector from the Turbomolecular Pump
Controller Sled.
d. Loosen by hand (or with a Phillips screwdriver) the fastener that
holds the Turbomolecular Pump Controller Sled to the tower.
e.
Remove the Turbomolecular Pump Controller Sled from the tower.