![Thermo Finnigan LCQ Deca XP Plus Hardware Manual Download Page 137](http://html.mh-extra.com/html/thermo-finnigan/lcq-deca-xp-plus/lcq-deca-xp-plus_hardware-manual_1104316137.webp)
Service Maintenance
____________________________________________________________ Replacing a Fuse
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-31
5.6 Replacing a Fuse
Fuses protect the various circuits by opening the circuits whenever
overcurrent occurs. On the MS detector, most of the fuses are located on the
System Control PCB. Several fuses, however, are located on the RF Voltage
Amplifier PCB, Analyzer Auxiliary PCB, and Power Module. The function
and current rating of the various fuses are listed in Table 5-5.
Check fuses when power is lost to a fused subsystem.
CAUTION. Always place the electronics service switch in the Service
Position (or shut down the system and disconnect the power cord) before
you replace fuses on the System Control PCB, RF Voltage Amplifier PCB,
or Analyzer Auxiliary PCB.
Table 5-5.
MS detector fuses
Location
Fuse
Circuit
Description
P/N
System
Control PCB
F1
Multipoles and tube
lens power supplies,
ion gauge grid
0.25 A, time lag, 5 x 20 mm, 250 V
00006-01700
System
Control PCB
F2
Multipoles and tube
lens power supplies
0.25 A, time lag, 5 x 20 mm, 250 V
00006-01700
System
Control PCB
F3
Ion transfer capillary,
multipoles, and mass
analyzer DC offsets
0.16 A, time lag, 5 x 20 mm, 250 V
00006-01700
System
Control PCB
F4
Ion transfer capillary,
multipoles, and mass
analyzer DC offsets
0.16 A, time lag, 5 x 20 mm, 250 V
00006-01700
System
Control PCB
F5
Ion gauge filament
3.15 A, time lag, 5 x 20 mm, 250 V
00006-10510
System
Control PCB
F6
Ion transfer capillary
heater
3.15 A, time lag, 5 x 20 mm, 250 V
00006-10510
System
Control PCB
F7
Conversion dynode
power supply
0.25 A, time lag, 5 x 20 mm, 250 V
00006-11204
System
Control PCB
F8
RF detector power
supply
0.4 A, time lag, 5 x 20 mm, 250 V
00006-05080
System
Control PCB
F9
RF detector power
supply
0.4 A, time lag, 5 x 20 mm, 250 V
00006-05080
System
Control PCB
F10
F11
APCI vaporizer
heater
2.5 A, time lag, 5 x 20 mm, 250 V
00006-09510
Summary of Contents for LCQ Deca XP Plus