Operation
SORVALL
Centrifuges
4-12
6. Press HOLD to select a continuous run.
7. Press TEMP and enter the desired rotor temperature.
8. Press START to begin rotor temperature equilibration. The cen-
trifuge will alert you that a sample overtemperature condition
exists, and that pre-cooling is taking place – press CLEAR.
9. Monitor progress of the value in the RUN TEMPERATURE
display, and note the time when the displayed RUN tempera-
ture reaches the SET temperature. From the time it does, allow
the run to continue for an additional 30 minutes, then press
STOP.
Running Hazardous Material
The
RC3
BP™ Centrifuge is designed with a refrigeration system
that operates independently of the laboratory environment. Be-
cause the centrifuge chamber is not designed for biocontainment,
some vapors or aerosols released from uncapped, leaking or broken
tubes may leak from the chamber during operation. Once a run is
completed and the chamber door is opened, the vapors or aerosols
which have concentrated in the chamber will be released into the
laboratory area. For this reason, when materials which are patho-
genic, toxic, or otherwise hazardous in nature are to be run, the
centrifuge should be located in a biohazard safety enclosure and
operated with all appropriate safety precautions. An added level of
protection can be attained through the use of SORVALL
®
biocon-
tainment rotors.
Read the WARNING regarding hazardous
materials found on the Safety Information Page in the
front of this manual.
Use appropriate decontamination procedures should exposure to
any hazardous material occur. See Chapter 5 for procedure to fol-
low if a centrifuge or rotor that has been used with a hazardous
material must be returned to our service facilities for repair.
W A R N I N G
If using pathogenic, toxic,
or otherwise hazardous materials in
the
RC3BP™, we recommend that
the centrifuge be located in a biohaz-
ard safety enclosure and operated
with all appropriate safety precau-
tions.
!