2-1
RC3 BP™
Installation
Chapter 2: INSTALLATION
This chapter contains instructions to prepare the SORVALL
®
RC3
BP™ centrifuge for operation.
Inspection
1. When you receive your centrifuge, carefully inspect it for any
signs of shipping damage. If you find damage, report it imme-
diately to the transportation company and file a damage claim,
then notify
Thermo
.
2. Check the parts received against the shipping list; if any parts
are missing, contact
Thermo
(see back cover).
3. Remove all packing material, and remove any remaining items
from inside the rotor chamber (where this manual was).
Electrical Requirements
The
RC3
BP™ has specific power requirements and must be con-
nected to the correct supply for proper performance. The name-
plate on the back panel of the cabinet specifies one of the following
AC supplies:
200 V, 50 Hz, 30 A, single Ø
200 V, 60 Hz, 30 A, single Ø
220 V, 50 Hz, 30 A, single Ø
208 V, 60 Hz, 30 A, single Ø
230 V, 50 Hz, 30 A, single Ø
220 V, 60 Hz, 30 A, single Ø
230 V, 50 Hz, 20 A/P, poly Ø
230 V, 60 Hz, 30 A, single Ø
240 V, 50 Hz, 30 A, single Ø
The supply voltage should be checked with a voltmeter, then you
should verify that the voltage indicated on the nameplate on the
back panel is in agreement with the measured line voltage.
If the measured line voltage is not within 10% of the voltage speci-
fied on the nameplate, do not connect the power cord and operate
the centrifuge or damage to the centrifuge may result. Contact
Thermo
or your local representative of SORVALL
®
products to see if
the centrifuge can be modified for your voltage. The centrifuge
system includes a ground wire with 30 A circuit protector on each
positive power lead.
!
C A U T I O N
The centrifuge can be
damaged if it is connected to the
wrong voltage. Check the voltage
listed on the nameplate, and measure
the voltage at the power source before
plugging in the power cord.
Thermo
is not responsible for incorrect
installation.