RC3 BP™
Operation
4-9
SELECT BRAKE OFF?
This screen allows you to select a speed (in rpm) at which, after
termination, deceleration will transition from braking to a coast-
ing stop. The amount of time it takes for the rotor to stop
depends on the transition speed, windage and inertia of the
rotor. BRAKE OFF transition speed is set independently of set
run speed, and is not affected by changes to set run speed. If the
transition speed is set higher than the set run speed, at run
termination, the centrifuge will coast to a stop from set speed.
After selection, CHANGE OVERTEMPERATURE ALERT? will
appear in the display.
NOTE
If SLOW STOP and BRAKE OFF are both selected:
– If the BRAKE OFF transition speed is set to 500 rpm
or higher, the SLOW STOP selection will be ignored.
– If the BRAKE OFF transition speed is set below 500
rpm, the centrifuge will decelerate with full braking to
500, transition to the specified SLOW STOP rate,
then change to a coasting stop when the specified
BRAKE OFF transition speed is reached.
CHANGE OVERTEMPERATURE LIMIT?
The
RC3
BP™ has sample overtemperature protection that will
terminate a run in progress if the calculated sample temperature
in the RUN display exceeds SET temperature by more than a
specified amount. CHANGE OVERTEMPERATURE LIMIT al-
lows you to specify a maximum allowable sample temperature
as it relates to the current SET temperature, to create a new
overtemperature offset value that will be applied to future
changes to set temperature. Overtemperature limit maximum
temperature cannot be set below 2
°
C over the SET temperature.
The offset value will be retained until overwritten. (See CHANGE
OVERTEMP LIMIT on page 3-9 for more information.)
The
RC3
BP™ also has an automatic precool feature, and over-
temperature termination is disabled when precooling is in
progress. The precool feature sounds an alarm and "SAMPLE
OVERTEMP – PRE-COOLING" appears in the SET display if
the calculated sample temperature in the RUN display is more
than 2
°
C above the SET temperature, either when START is
pressed, or during a STEP RUN, when a new program's SET
temperature is more than 2
°
C lower than the RUN temperature.
When CLEAR is pressed, the DEG C field in the SET display
will change to P-COOL, which will remain in the display until
the RUN temperature comes into range, or the run terminates.
After selection, SELECT STEP RUN? will appear in the display.