Operation
SORVALL
Centrifuges
4-2
Rotor Installation, Loading and
Balancing
Install, load, and balance the rotor according to the instructions
given in the rotor instruction manual.
W A R N I N G
If maximum compartment
mass is exceeded, the maximum
speed must be lowered (see Reduc-
ing Speed for Rotor Compartment
Loads in Excess of Design Mass,
page 4-13). Failure to do so can cause
rotor failure which could result in per-
sonal injury or centrifuge damage.
!
Table 4-1. ROTOR INFORMATION
Maximum
Maximum
Maximum
Balance
Critical
Biohazard
Speed
RCF
Compartment
Margin
Speed
Containment
Rotor
(rpm)
(
g-force)
1
Mass (grams)
2
(grams)
(rpm)
Available
SWINGING BUCKET
H-6000A
5000
7277
2575
3
10
450
yes
HBB-6
5000
7129
2800
3
10
450
no
HLR-6
5000
7205
2800
3
10
450
no
H-6000A M
5000
5186
462
10
450
no
(with microplates)
H-2000B
2800
2297
750
10
800
yes
H-2000B M
2800
1822
290
10
800
no
(with microplates)
HL-2B
3300
2024
2000
3
5
700
no
HBB-12
2900
2462
3820
3
25
500
no
H-4000
5250
7121
2575
3
10
450
no
FIXED-ANGLE
LA/S-400
6000
5646
90
10
1000
yes
LAC-6000
6000
7890
1500
25
450
no
1 Values reflect the centrifuge's calculation rounding, are based on the rotor's maximum radius, and do not consider any tube, bottle or adapter
geometry.
2 Maximum allowable at maximum speed (see Reducing Speed for Rotor Compartment Loads in Excess of Design Mass, page 4-13).
3 Value includes the mass of the bucket or carrier.