Bench Status indicator
To the right of the Experiment drop-down list box is the Bench
Status indicator.
If the indicator is a green check mark, the spectrometer has passed
all of its diagnostic tests.
If the indicator is a yellow circle, a cooled detector in the
spectrometer has become warm. A message appears explaining the
problem and allowing you to access information about correcting it.
If the indicator is a red X, the spectrometer has failed a diagnostic
test and requires corrective action. A message appears explaining the
problem and allowing you to access information about correcting it.
Note
If you want to view the on-line videos to help troubleshoot a
problem, put the spectrometer’s Tutorial CD into the computer’s
CD-ROM drive.
S
Toolbar
At the left side of the window is the toolbar. (You can use Edit
Toolbar in the Edit menu to position the toolbar at the top or right
side of the OMNIC window as well. Each button in the toolbar
represents a command or other feature. To see the name of the
command or feature, point to the button and wait a moment. To
initiate the command or feature, click the button with the left mouse
button.
?
In OMNIC Help Topics find “toolbar” in the Index and go to
“Displaying and using a toolbar” for information about using and
customizing the toolbar buttons.
26 Thermo Electron Corporation