Problem
Possible Cause
Solution
The Scan status
indicator does not flash;
the system is not
scanning.
Note:
After 1 hour
without data collection
activity, the
interferometer stops
scanning and the Scan
light stays on. This is
normal. Any data
collection activity will
cause the interferometer
to begin scanning.
The spectrometer is turned off.
The spectroscopy software has
not been started.
The spectrometer is out of
alignment.
The electronics module needs to
be reset.
The laser is flickering.
The laser is off.
Continued on next page...
Turn on the spectrometer power.
Start OMNIC (or other software you
are using).
Align the spectrometer using the
Align button on the Diagnostic tab
of the Experiment Setup dialog box.
Turn off the power to the computer
and the spectrometer. Leave the
power off for 15 seconds. Turn on
the spectrometer and wait until the
start-up diagnostics finish. Then turn
on the computer and restart OMNIC.
Click the laser indicator on the
Diagnostic tab of the Experiment
Setup dialog box to check the laser
operation. Call Thermo Electron for
assistance.
Turn off the spectrometer power,
unplug the power cord, remove the
spectrometer cover and make sure
the laser cable is seated properly.
See “Replacing the laser” in the
“Servicing your spectrometer” book
in Spectrometer Help Topics for
instructions.
Warning:
Never open
the spectrometer cover without first
turning off the power and
unplugging the power cord.
Nicolet FT-IR User’s Guide 145