Problem Possible
Cause Solution
After data collection, a
message says the
sample and
background spectra
have the wrong
resolution.
The resolution of the current
background spectrum does not
match the resolution selected
for the sample spectrum
collection.
Change the Resolution to match the
background and collect the spectrum again.
Use the Resolution parameter on the Collect
tab of the Experiment Setup dialog box.
An experiment
provided with OMNIC
does not produce good
results.
The parameter settings in the
experiment file have been
changed so that the experiment
is no longer useful.
Restore the experiment to its default
settings: First open the experiment using
Experiment Setup and save it using a new
filename if you do not want to overwrite it.
Then open the experiment in the
FACTORY directory whose filename is
same as the original experiment and save it
in the OMNIC\PARAM directory.
The desired
experiment does not
appear in the
Experiment drop-
down list box.
The experiment was installed
with the software but has not
yet been opened.
If the experiment was designed for a Smart
Accessory, install the accessory; the
experiment will be opened automatically
and will appear in the Experiment drop-
down list box. If the experiment was not
designed for a Smart Accessory, open it
using Experiment Setup; it will be added to
the list.
Libraries created with
previous version of
OMNIC are not listed
in the Library Setup
dialog box.
The libraries are not in one of
directories listed in the dialog
box (typically LIBS within the
OMNIC directory).
Use the Add Directory button on the Search
Libraries tab to add the directory.