Follow these steps to perform the performance test:
1. Make sure the main sample compartment is empty.
Remove any samples and sampling accessories (except a
transmission sample holder).
2. Exit
OMNIC.
This ensures that the Bench Diagnostics software will run
correctly.
3. Start the Bench Diagnostics software.
To do this, click the Start button on the Windows taskbar, point
to Programs (All Programs in Windows XP), point to the
Thermo Nicolet folder and then choose Bench Diagnostics.
See “Performing advanced diagnostic tests” in the
“Troubleshooting Hardware Problems” chapter for more
information about running the diagnostics software.
4. Click the Performance Tests button and then follow the
instructions that appear on the screen.
Note
If the noise readings are greater than normal or have changed
significantly, call the Thermo Electron Technical Support for
assistance. Outside the U.S.A., call your local service representative.
Telephone numbers for all Thermo Electron offices are provided
with your system.
S
If special performance tests are available for your spectrometer,
a button labeled “Extended Tests” (or another appropriate name)
will appear. Click this button and follow the instructions that
appear on the screen to perform the tests.
Nicolet FT-IR User’s Guide 133