View Modes
While in the operational display, pressing the button allows
you to view but not change the following parameters:
• Channel 1.
• Channel 3.
• Setpoint 1.
• Setpoint 3.
• Setpoint 5.
• Peak (of CH1).
• Total 1 (total of CH1).
While in the operational display, pressing the button allows
you to view but not change the following parameters:
• Channel 2.
• Channel 4.
• Setpoint 2.
• Setpoint 4.
• Setpoint 6.
• Valley (of CH1).
• Total 2 (total of CH2).
SP1 SP2 SP3 SP4 SP5 SP6
Prog.
Operational Display
Press
4 secs
SP1 SP2 SP3 SP4 SP5 SP6
Prog.
Operational Display
Press
1
After configuring an on demand
function in the
Calibration On
Demand Mode
, press the
P
button for 4 seconds to
activate one of the following
selected on demand modes.
Tare
Single-point calibration
Two-point calibration
Primary input compensation
Manual loader (manual offset)
CH3
SP_1
SP_3
SP_5
PEAK
TOT_1
CH1
To view, press
the button:
SP1 SP2 SP3 SP4 SP5 SP6
Prog.
Operational Display
Press
1
View Mode
CH4
SP_2
SP_4
SP_6
VALEY
TOT_2
CH2
To view, press
the
button:
On Demand Modes
The meter can be programmed to activate the following func-
tions on demand by pressing the
P
button for 4 seconds:
• Tare.
• Single-point calibration.
• Two-point calibration.
• Primary input compensation.
• Manual loader (manual offset).
• Print.
The on demand function is selected in the calibration mode.
OR
OR
OR
OR
OR
P
SP1 SP2 SP3 SP4 SP5 SP6
Prog.
Operational Display
SP1 SP2 SP3 SP4 SP5 SP6
Prog.
Operational Display
P
SP1 SP2 SP3 SP4 SP5 SP6
Prog.
Operational Display
P
On Demand Modes
Front Panel Programming Codes continued
For a full breakdown of all programming codes, see the
Tiger 320 Series Programming Code Sheet (NZ101).
View Mode