5. FAQ
5.1 The tester will not turn on.
Ensure the power cord is connected. Ensure the line input is set to the correct line
voltage. Check to make sure the fuse is not blown. See page 93.
5.2 The panel keys are not working.
Ensure the tester is not in remote mode, page 66.
Ensure the tester is not in SIGNAL I/O or Remote Connect mode, page 57.
5.3 When I press the START button the tester will not start testing?
The tester must first be in the READY status before a test can be started. Ensure
the tester displays READY before pressing the START button, page 32 (manual
test), 49(automatic test).
If “Double Action” is enabled, the START button must be pressed 0.5 seconds after
the STOP button is pressed, otherwise the tester will not start testing.
If “Interlock” is enabled, the interlock key must be inserted into the signal I/O port
on the rear before a test can be started. See page 62 for details.
Lastly, ensure that the Start Ctrl setting is correctly configured in the Common
Utility menu. For example, to enable the START button to start a test, ensure that
the Start Ctrl setting is set to FRONT PANEL. See page 58 for details.
5.4 The accuracy does not match the specification.
Make sure the tester is powered on for at least 30 minutes, 15°C~+35°C.
This is necessary to stabilize the unit to match the specification.
For more information, contact your local dealer or TEXIO at www.texio.co.jp/