28
Steps
1.
Press the UTILITY key on the front
panel when the tester is in MANU/EDIT
status. The display will go from the
normal EDIT status to the MANU Utility
menu for the current test.
UTILITY
L I T Y
I
M A U
N
U T
2
M A N U = * * * - 0 0
A R C M D E : O
O
F
F
G R O U N D M O D E :
N
O
P A S S H
L D :
O
O F
F
F A I L M D E :
O
P
S T O
M A X H O L D : O F F
2.
Use the UP/DOWN arrow keys to move
to the FAIL MODE setting.
3.
Use the scroll wheel to set FAIL MODE.
FAIL MODE CONTINUE, HOLD, STOP
4.
Press the EDIT/SAVE key to save and
exit the MANU Utility menu.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the MANU
Utility menu to cancel and exit.
2.3.14 Setting MAX HOLD
Background
The MAX HOLD setting will hold the maximum current
measured in the ACW and DCW tests or the maximum
resistance measured in IR and GB tests.
Steps
1.
Press the UTILITY key on the front
panel when the tester is in EDIT status.
The display will go from the normal
EDIT status to the MANU Utility menu
for
the current test
.
UTILITY
L I T Y
I
M A U
N
U T
2
M A N U = * * * - 0 0
A R C M D E : O
O
F
F
G R O U N D M O
P A S S H
L D :
O
O F
F
F A I L M D E :
O
P
S T O
M A X H O L D :
O F F
D E :
N
O
Note
The MANU UTILITY settings only apply to the selected
MANU test.
2.
Use the UP/DOWN arrow keys to move
to the MAX HOLD setting.