21
Steps
1.
Press the HI/LO soft-key or use the UP
/ DOWN arrow keys to bring the cursor
to the HI SET (ACW/DCW/GB) setting
or the LO SET (IR) setting.
I / L
H
O
OR
I R
G B
A CW
D CW
I / L
H
O
I M E
T
R
E
F R E Q =
0 H z
6
0
100
k
V
I
D
E
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
I
H
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
T
2
M A N U = * * * - 0 0
cursor
m A
2.
Use the scroll wheel to set the HI
SET/LO SET limit.
HI SET
STW-9900
STW-9800
ACW (HI)
0.001mA~110.0mA
0.001mA~042.0mA
DCW (HI)
0.001mA~021.0mA
0.001mA~011.0mA
IR (LO)
0.001G
Ω ~ 50.00GΩ
0001MΩ ~ 9999MΩ
GB (HI)
000.1mΩ ~ 650.0mΩ
-
3.
Press the HI/LO soft-key again or press
the DOWN arrow key to switch
between HI SET and LO SET.
I / L
H
O
OR
I R
G B
A CW
D CW
I / L
H
O
I M E
T
R
E
F R E Q =
0 H z
6
0
100
k
V
I
D
E
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
0
L
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
T
2
M A N U = * * * - 0 0
cursor
1 . 0 0 m A
I
H
E
S
= 0
T
m A
4.
Use the scroll wheel to set the HI
SET/LO SET limit.
LO SET
STW-9900
STW-9800
ACW (LO) 0.000mA~109.9mA
0.001mA~042.0mA
DCW (LO) 0.000mA~020.9mA
0.001mA~011.0mA
IR (HI)
0.001GΩ~50.00GΩ
0001MΩ ~ 9999MΩ
GB (LO)
000.0mΩ ~ 649.9mΩ
-
Note
The LO SET setting is limited by the HI SET setting. The
LO SET limit cannot be greater than the HI SET limit.
When setting the current, be aware that a maximum of