81
Parameter/ Return
parameter
<NR2>
0.000 ~ 50.00
0 ~ 9999
(STW-99xx :
GΩ)
(STW-98xx :
MΩ)
Example
MANU:IR:REF 0.900
Sets the IR reference to 900 MΩ.
4.8.23 MANU:GB:CURRent
Set
Query
Description
Sets or returns the GB current in A. The test must first be in
GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:CURRent <NR2>
MANU:GB:CURRent?
Parameter/ Return
parameter
<NR2>
3.00~33.00
Example
MANU:GB:CURR 3.00
Sets the GB current to 3.00A.
4.8.24 MANU:GB:RHISet
Set
Query
Description
Sets or returns the GB HI SET resistance value in mΩ. The
test must first be in GB mode before this command can be
used.
Syntax
Query Syntax
MANU:GB:RHISet <NR2>
MANU:GB:RHISet?
Parameter/ Return
parameter
<NR2>
000.1 ~ 650.0
Example
MANU:GB:RHIS 100.0
Sets the HI SET value to 100mΩ.
Note
If the (GB current x HI SET resistance) > 5.4V, then an error
will be generated (“GBV > 5.4V”).
4.8.25 MANU:GB:RLOSet
Set
Query
Description
Sets or returns the GB LO SET resistance value in mΩ. The
LO SET value must be less than the HI SET value. The test
must first be in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/ Return
parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50mΩ.
4.8.26 MANU:GB:TTIMe
Set
Query
Description
Sets or returns the GB test time in seconds. The test must first
be in GB mode before this command can be used.