79
4.8.16 MANU:DCW:REF
Set
Query
Description
Sets or returns the DCW reference value in mA. The test must
first be in DCW mode before this command can be used.
The reference value must be less than the HI SET value.
The reference value uses the same range as the HI SET
value.
Syntax
Query Syntax
MANU:DCW:REF <NR2>
MANU:DCW:REF?
Parameter/ Return
parameter
<NR2>
0.000 ~ 20.9
Example
MANU:DCW:REF 0.01
Sets the DCW reference to 0.01 mA.
4.8.17 MANU:DCW:ARCCurrent
Set
Query
Description
Sets or returns the DCW ARC current value in mA. ARC must
be enabled to set the ARC current. The test must first be in
DCW mode before this command can be used.
ARC current uses the same range as the HI SET value. The
ARC current is limited to 2X the HI SET value.
Syntax
Query Syntax
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
Parameter/ Return
parameter
<NR2>
2.000 ~ 40.0
Example
MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.
4.8.18 MANU:IR:VOLTage
Set
Query
Description
Sets or returns the IR voltage in kV. The test must first be in IR
mode before this command can be used.
Syntax
Query Syntax
MANU:IR:VOLTage <NR2>
MANU:IR:VOLTage?
Parameter/ Return
parameter
<NR2>
0.05 ~ 1 (0.05kV to 1kV: steps of .05)
Example
MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.