80
4.8.19 MANU:IR:RHISet
Set
Query
Description
Sets or returns the IR HI SET resistance value. The test must
first be in IR mode before this command can be used. Unit of
STW-
98xx is MΩ, and of STW-99xx is KΩ.
Syntax
Query Syntax
MANU:IR:RHISet <NR2>|NULL
MANU:IR:RHISet?
Parameter/ Return
parameter
<NR2>
0.002 ~ 50.00
2 ~ 9999
(STW-99xx :
GΩ)
(STW-98xx :
MΩ)
NULL
Sets the HI SET value to high impedance
Example
MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 MΩ.
4.8.20 MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value. The LO SET
value must be less than the HI SET value. The test must first
be in IR mode before this command can be used. Unit of
STW-
98xx is MΩ, and of STW-99xx is KΩ.
Syntax
Query Syntax
MANU:IR:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/ Return
parameter
<NR2>
0.001 ~ 50.00
1 ~ 9999
(STW-99xx :
GΩ)
(STW-98xx :
MΩ)
Example
MANU:IR:RLOS 0.010
Sets the IR LO SET resistance to 10MΩ.
4.8.21 MANU:IR:TTIMe
Set
Query
Description
Sets or returns the IR test time in seconds. The test must first
be in IR mode before this command can be used.
Syntax
Query Syntax
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
Parameter/ Return
parameter
<NR2>
1.0 ~ 999.9 seconds
Example
MANU:IR:TTIM 1
Sets the IR test time to 1 second.
4.8.22 MANU:IR:REF
Set
Query
Description
Sets or returns the IR reference value. The test must first be
in IR mode before this command can be used. The reference
value must be lower than the HI SET value. Unit of STW-98xx
is MΩ, and of STW-99xx is KΩ.
Syntax
Query Syntax
MANU:IR:REF <NR2>
MANU:IR:REF?