32
2.3.16 Saving and Exiting EDIT Status
Background
After all test parameters have been set, the test can be
saved. After a test is saved it can be used when creating
an AUTO test.
Warning
The special MANU number, 000, can be saved, however it
cannot be used for AUTO tests. See page 39 for details.
Steps
1.
When in EDIT status, press the
EDIT/SAVE key to save the current
test. This will enter the VIEW status for
the chosen test number.
EDIT/SAVE
I R
G B
A CW
D CW
E
F R E Q =
0 H z
6
0
100
k
V
E
I
V
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
I
H
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
W
2
M A N U = * * * -
0 0
I
D
E
T
m A
2.
The Status changes from EDIT to VIEW.
Note
Pressing the EDIT/SAVE key again will return the tester
back to EDIT status for the current test.
2.3.17 Running a MANU Test
Background
A test can be run when the tester is in READY status.
Note
The tester cannot start to run a test under the following
conditions:
A protection setting has been tripped; when a protection
setting has been tripped the corresponding error
message is displayed on the screen. See page 94 for a
comprehensive list of the all the setting errors.
The INTERLOCK function is ON and the Interlock key is
not inserted in the signal I/O port (page 58).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START button is
pressed immediately after the STOP button (<0.5s).
Note
When a test is running the voltage output cannot be
changed, unless the test is under the special manual
mode. See page 39 for details.
Steps
1.
Ensure the tester is in VIEW status for
the current test. Save the current test if
necessary.
Page 32