Calibrate the ADC Device on the EVM
12
SLAU776 – May 2018
Copyright © 2018, Texas Instruments Incorporated
Setup Procedure
2.10 Calibrate the ADC Device on the EVM
shows the GUI open to the
Control
tab.
Figure 2-3. Configuration GUI: Control Tab
1. With the EVM GUI open on the PC, navigate to the
Control
tab.
2. To calibrate the ADC, click the
Cal Triggered/Running
button once, and then click the button again.
This action stops and starts the calibration engine.
NOTE:
The
Cal Triggered/Running
button executes a calibration sequence that is required for full
performance. This calibration is performed automatically during the previous step shown in
, but must be performed again any time the sampling rate changes, after
significant temperature change of the ADC, or after exiting the power-down mode. See the
for details regarding the necessary calibration sequence.
3. Select
Foreground
,
Background
, or
Low Power Background
calibration mode using the
Sampling and
Calibration Mode
drop-down menu on the
EVM
tab.