1.8 Connection to the power supply
Danger
The machine must be connected to the building power system by
a qualified electrician.
In accordance with the pertinent standards, the yellow-green con-
nection terminal must be attached to the earthing system before
additional electrical connections are made.
Before making the electrical connections, please study the en-
closed wiring diagram. Also check the machine rating plate to
make sure that the ratings of the building power supply conform
to the requirements for voltage, wattage, amperage, and fre-
quency of the machine.
The electrical socket must have a safety device that will protect
the system against over-current. This safety device must satisfy
the stipulations of the relevant standards, and must match the ma-
chine voltage. The technical characteristics of this safety device
must also satisfy the standards that apply in the country in which
the machine is installed.
Caution
The manufacturer of the machine cannot be held liable for any dam-
ages that result because the information here is not observed.
Electrical tolerances:
Actual voltage:
10% of the rated voltage
Frequency:
1% of the rated frequency, continually;
2% of the rated frequency, on a
short-term basis
The manufacturer shall not be liable for damages to persons or property that arise be-
cause the above instructions have not been observed.