Instruction Manual for TESA TWIN-SURF
13
6
FUNCTIONS
6.1 Calibration
To perform instrument calibration, enter main menu and select the item
Calibration
through
the key
. The following screen appear.
The steps are as follows:
1.
Enter nominal value
: Move with the button
to the box with the numerical
value and change the value by pressing
until the desired value is obtained.
2.
Start calibration
: when the desired value is displayed, press
to go to the
start measure
item, then press
to start the calibration.
3. At the end of the measurement, the message
calibration done
is shortly displayed.
No data is now available on the main menu. Perform a new measurement to display
the last measuring value.
Calibration with Ra if ISO 4287 standard
Calibration with R if ISO 12085 standard
Calibration based on a measurement average is available from the software
TESA DATA-STUDIO with max 5 measurement for a calibration.
6.2 Parameters
To access the parameters selection screen, enter the main menu and select the
Parameters
item through the key
. The following screen appear.