Accuracy Specifications (3)
Overload Protection:
Glass fuse 0.5A, 250V, fast type, 5 x 20mm.
H. Temperature (Model UT60C)
Range
Resolution
Accuracy
-40°C to 1000°C
1°C
-40°C to 0°C
±(3% + 4)
0°C to 400°C
±(1% + 3)
400°C to 1000°C
±(2% + 10)
I. DC Current
Range
Resolution
Accuracy
Overload Protection
400
µ
A
0.1
µ
A
±(1% + 2)
0.5A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
4000
µ
A
1
µ
A
40mA
0.01mA
±(1.2% + 3)
400mA
0.1mA
4A
0.001A
±(1.5% + 5)
10A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
10A
0.01A
Remarks:
•
4A and 10A Range:
For continuous measurement
≤≤
10 seconds and interval not less than 15 minutes.
Accuracy Specifications (4)
J. AC Current
Range
Resolution
Accuracy
Overload Protection
400
µ
A
0.1
µ
A
±(1.5% + 5)
0.5A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
4000
µ
A
1
µ
A
40mA
0.01mA
±(2% + 5)
400mA
0.1mA
4A
0.001A
±(2.5% + 5)
10A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
10A
0.01A
Remarks:
Frequency response 40Hz to 400Hz.
UT60C
displays effective value of since wave (mean value response) displays true rms
value.
•
4A and 10A Range:
For continuous measurement
≤≤