7.2.4 -
Image Setting
In Image Menu, press the icon ″ ″ is how the page of Image
Setting. It contains the ″ISO″ and “Clear Screen. ″ISO″ is used
for isothermal analysis. It allows to Thermal Imagers displays
a setting colour for Infrared Image, which temperature is in the
given range.
Press ″ISO″ bar to switch on/off isothermal analysis. The Icon
″ ″ means to open isothermal analysis; the Icon ″ ″ means
to close isothermal analysis. Press the icon ″ ″ popup the
submenu of ″ISO″ settings. This submenu contains settings
for Isothermal mode, Isothermal temperature, Isothermal
range and Isothermal colour. Isothermal mode has Inter,
Above, Below. The Inter mode means infrared image which
the temperature in the range of [Level-Width/2, Level+Width/2]
is set to Colour; The Above mode means the infrared image
which the temperature is greater than Level+Width/2 is set to
Colour; The Below mode means the infrared image which the
temperature is less than Level-Width/2 is set to Colour. The
Colour has Black, White, Green and Red to select.
Press ″Clear Screen″ bar to show only image in the screen.
The Icon ″
″ means to clear screen, The Icon ″ ″ means
to don’t clear screen.