36
Sidekick
®
Plus Test Set
Water
1. Press
F4 More Cable
until the reflection is
displayed.
2. Use the up (
2
) and down (
8
) arrows keys to adjust
the waveform height.
3. Use the left (
4
) and right (
6
) arrows to move the
cursor to the beginning of the wet section (“A” on
the illustration below). This is the distance to the
water.
4. Use the left (
4
) and right (
6
) arrows to move the
cursor to the end of the wet section (“B” on the
illustration below).
“A” to “B” is the wet section.
Note: The distance from the front panel to the water
(A) is correct. The wet section distance (A to B) is not
correct because water changes the velocity of propaga-
tion (Vp). Subtract the dry section distance from the
cable map to obtain the wet distance. Or measure from
both ends of the cable to the wet section.
Comparing Two Pairs
The Sidekick
®
Plus test set can display two TDR traces
at the same time or perform mathematical comparisons
of the traces. The accessory kit contains a second set
of reference leads (yellow and blue).
1. Connect the primary test leads to the first pair:
red to ring and black to tip.
2. Connect the yellow and blue reference leads to the
second pair.
3. Move the rotary knob to the TDR position.
4. Press
F3 Config
.
5. Press the down (
8
) key until “Test Type” is
highlighted.