![tell gsm easyaid Installation And Application Manual Download Page 16](http://html1.mh-extra.com/html/tell/gsm-easyaid/gsm-easyaid_installation-and-application-manual_1083889016.webp)
16
Miscellaneous options:
o
Low battery warning:
if enabled, the device plays a warning voice message by
15 minutes when the battery voltage drops below 7.5V on power loss, until the
battery is discharged and the device turns off, or the power voltage restores.
o
Number of rings:
it is possible to specify, after how many rings (1 to 5) the
module should answer a call automatically.
o
Ring volume for incoming calls:
you can adjust the ringer volume for incoming
calls between 50% and 100%.
o
RF transmitter supervision timeout:
the RF transmitter supervision interval
can be set between 1 and 24 hours. If the device does not receive the test signal
from the RF transmitter within this interval, then it generates “
RF supervision
fault
” event.
Nurse code:
a code consisting of digits 1,2,3 and 4 can be specified which can be
used by the social workers/nurses to confirm a visit.
The preset code can be entered
on device using the 1-4 function keys. After the social worker/nurse enters the code,
the module generates
“
Nurse confirmation
” event and initiates call and/or sends SMS
as configured.
Reminder settings:
four daily reminders can be configured for diverse tasks, as
specified in the “
User’s guide
” chapter.
o
Number of reminder repetitions:
it can be configured between 1 and 10 how
many times the module should repeat the given reminder voice message.
o
Delay between repetitions:
it can be configured between 5 and 30 seconds,
how long the module should wait before repeating the given reminder voice
message.
o
1-4:
the checkboxes can be used to enable the 1-4 reminders, and in the time
fields the time of day can be set when the module should play the given reminder
voice message.