![tell gsm easyaid Installation And Application Manual Download Page 10](http://html1.mh-extra.com/html/tell/gsm-easyaid/gsm-easyaid_installation-and-application-manual_1083889010.webp)
10
The battery of the RF transmitter needs to be replaced forthwith if
“
RF transmitter
battery low
” event is received, as well as at least yearly, independently of the reception
of this event, in order to retain the operating range of the transmitter! The user can
initiate emergency alarms by 5 seconds using the RF transmitter, while from the
easyAID unit this can be done at anytime.
To cancel or revoke an emergency call press the Cancel
key while the countdown
is shown on the digital display. After an emergency alarm is initiated, the device waits 5
seconds before initiating the call and sending SMS message, which time is available to
cancel an alarm initiated by mistake.
When an emergency call is initiated, the device calls the 1-4 phone numbers in order,
and then sends the configured SMS messages. If any of the called numbers/party
received and confirmed the call, then the device does not initiate further calls to the next
phone numbers and does not send any SMS messages about this event.
ATTENTION! Disable voicemail service on the phone numbers to be dialed by the
device because in case that the given phone number is not available, this causes
significant delay to the calls and SMS messages to be initiated to the next phone
numbers! The outgoing calls which are received by voicemail are considered
successful by the device!
Confirming an emergency call
The called party can confirm the emergency call by pressing
key on the phone. If the
called party receives the call and presses
key on the phone, the device will not initiate
further calls to the next phone numbers, considers the emergency call process finished,
therefore will not send any configured SMS messages about this event.
2.4.5 Confirming a visit
The social worker can confirm the visit by entering a 4 digit individual password (nurse
code) consisting of digits 1,2,3 and 4, which has to be previously configured with the
programming software. If the password has been entered correctly, the module
generates
“
Nurse confirmation
” event and initiates the calls and/or SMS sending
configured for this event.
Entering the confirmation password:
1) Keep the Cancel
key pressed for at least 5 seconds
2) On digital display
„- - - -” is displayed
3) Enter your 4 digit nurse code using the 1-4 function keys
4) If the password has been entered correctly
„
ok
” is displayed
5) In case of wrong password
„
Err
” is displayed