TIENet® Model 310 Ex Ultrasonic Level Sensor
Section 2 Installation and Setup for Signature
2-16
2.5.2 Measurement Setup
From Measurement Setup (Figure 2-13 below), select Level Input
Setup to define the measurement range.
The
Minimum Blanking Distance
is the shortest distance
from the sensor face (highest expected liquid level). Because of
the dead band, this value can never be less than one foot.
Depending on the elevation of your sensor, this value may be
increased to help ensure that echoes read by the flow meter come
only from the surface of the flow stream, and not off the walls or
sides of the channel.
The
Maximum Blanking Distance
is the distance between the
sensor face and the bottom of the channel, or zero level. You can
enter a slightly larger value than calculated, if you prefer.
Figure 2-13 Configuring ultrasonic level measurement
Configure Options
Level Input Setup
Measurement Setup
1. Level Input Setup
2. Flow Input Setup
3. Volume Input Setup
1. 310 Level
310 Level
LEVEL
ft
ft
Blanking Distance
Minimum:
Blanking Distance
Maximum:
0
3
1. Site Setup
2. Measurement Setup
3. Adjust
4. Equation/Trigger Setup
5. Data Storage/Push Setup
6. Sampler Setup
7. Inputs/Outputs/Alarms Setup
8. Reset Totalizers
9. Reports/History Setup
1.
Press Next 2x.
Distance from sensor face to
maximum expected liquid
level
Distance from sensor face to
bottom of channel
Wall Mount
Approved for classified
hazardous locations.
Summary of Contents for TIENet 310 Ex
Page 2: ......
Page 4: ......
Page 14: ...TIENet Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction 1 6...
Page 35: ...TIENet Model 310 Ex Ultrasonic Level Sensor A 3...
Page 36: ...TIENet Model 310 Ex Ultrasonic Level Sensor A 4...
Page 38: ......
Page 40: ......
Page 42: ......