TIENet® Model 310 Ex Ultrasonic Level Sensor
Section 2 Installation and Setup for Signature
2-3
1. Remove one of the 6-position plug-in terminal strip connec-
tors from the case board.
Figure 2-1 TIENet Device terminal strips
2. If using a cord-grip fitting, install the cable nut in the
appropriate opening on the bottom of the Signature enclo-
sure, securing it to the wall with the lock nut (concave side
facing wall).
3. Feed the TIENet device cable end through the sealing nut
and seal, and through the cable nut. Lightly tighten the
sealing nut, just enough to hold the cable in place while
installing the connector.
Figure 2-2 Installing cable with a cord-grip fitting
4. Attach the wire ends to the terminal strip as shown in
Figure 2-3, then press the terminal strip back down into its
socket on the case board, as shown in Figure 2-4, taking
care not to strain any wire connections. Gently tug each
wire when finished, to verify secure connection to the
screw terminals.
Lock Nut
(concave side
facing wall)
Cable Nut
Seal
(color may vary)
Sealing Nut