TIENet® Model 310 Ex Ultrasonic Level Sensor
Section 2 Installation and Setup for Signature
2-14
2.5 Configuring the
System
To configure the Signature flow meter for operation with the
TIENet 310 Ex device, press
MENU
(
) to access the top
menu, and select Hardware Setup. For all TIENet devices
including the 310 Ex, select Smart Sensor Setup (TIENet).
2.5.1 Updating the Device
List
When the 310 Ex is physically added to the system, select
Perform Scan so that the flow meter detects it. When the scan is
complete, the 310 Ex appears in the list of connected devices,
ready to be configured with the steps shown in Figure 2-12 on the
following page.
Note
From the Hardware Setup menu, “Configure” refers to defining
and selecting the parameters for each connected device.
The four parameters that will appear for the 310 Ex device are:
310 Distance –
Distance between the bottom of the sensor and
the surface of the flow stream.
310 Air Temperature –
Temperature of surrounding (ambient)
air
310 Level –
Level of the flow stream surface
310 Signal –
Strength of the return echo
The name of any parameter can be customized by highlighting it
and pressing Enter (
) to display the character grid. Nav-
igate the grid using the arrow keys. Select characters with Enter
and clear characters with Delete (
).
Figure 2-11 Character grid
Done
Cancel
3 1 0 D i s t a n c e
A
B C D E F G H I J K L M N
b
a
Z
Y
X
W
V
U
T
S
R
Q
P
O
c d e f g h i
j k l m n o p
!
:
/
z
y
x
w
v
u
t
s
r
q
@ # $ % ^ & *
(
) -
_
+ = <
> ? ,
.
Summary of Contents for TIENet 310 Ex
Page 2: ......
Page 4: ......
Page 14: ...TIENet Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction 1 6...
Page 35: ...TIENet Model 310 Ex Ultrasonic Level Sensor A 3...
Page 36: ...TIENet Model 310 Ex Ultrasonic Level Sensor A 4...
Page 38: ......
Page 40: ......
Page 42: ......