10.2 Trigger Response Modes
NIRvana HS’s Trigger Response modes for Full Frame operation are:
•
No Response;
•
Start On Single Trigger;
•
Readout Per Trigger;
•
Expose During Trigger Pulse.
Timing mode, when combined with the Shutter Mode options, provide the widest
variety of modes for precision experiment synchronization.
10.2.1
External Trigger Input
The selected Trigger Response determines how the camera will respond to an External
Sync pulse supplied to the camera by the TRIGGER IN connector on the rear of the
camera. See
Figure 10-1: NIRvana HS Camera: Rear Panel
Things to keep in mind when setting up the External Sync pulse input include:
•
Pulse Height: 0 to +3.3 VDC logic levels (TTL-compatible.)
•
Pulse Width (trigger edge frequency): The time between trigger edges.
•
TRIGGER IN Connector Impedance: High impedance (~5 k
pull-up to +3.3 V.)
•
Trigger Polarity: Rising (+) or falling (-) must be indicated on the Trigger
expander.
4
4
11
-01
69
_0
028
Summary of Contents for Princeton Instruments NIRvana HS
Page 1: ...www princetoninstruments com NIRvana HS System Manual 4411 0169 Issue 1 0 January 8 2020 ...
Page 3: ...Issue 1 0 NIRvana HS System Manual 3 This page is intentionally blank ...
Page 4: ...4 NIRvana HS System Manual Issue 1 0 ...
Page 10: ...10 NIRvana HS System Manual Issue 1 0 ...
Page 26: ...26 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 34: ...34 NIRvana HS System Manual Issue 1 0 ...
Page 52: ...52 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 72: ...72 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 92: ...92 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 101: ...This page is intentionally blank ...
Page 106: ......