30
NIRvana HS System Manual
Issue 1.0
Figure 5-1: Adjustable C-Mount Adapter
2. Using a spanner wrench (or equivalent,) rotate the ring to the desired height.
NOTE:
The distance between the two adapter holes is 3.85”
[97.8 mm].
3. Use the hex wrench to tighten the setscrew to lock the adapter in place.
NOTE:
To lock the setscrew, the face of the adapter should be no
more than 0.1" (2.5 mm) out from the front surface of the
camera nose.
CAUTION!
!
The C-mount lens thread-depth should be 0.21" (5.33 mm)
or less. Otherwise, depending on the adapter in-out
location, the lens could bottom out and damage the glass. If
you are not certain of the thread depth, remove the adapter
from the camera nose, thread the lens into the adapter until
the lens threads are flush with the back surface of the
adapter. Note the depth at the front surface, remove the
lens, and then re-insert the adapter into the camera nose.
44
11-
0
16
9
_0
01
2
A
DJUSTABLE
D
UST
C
OVER
S
PANNER
H
OLES
(2) S
PANNER
W
RENCH
H
OLES
0.10” [2.5
MM
]
DIAMETER
X
0.15” [3.8
MM
] D
EEP
S
ET
S
CREW
C-M
OUNT
A
DAPTER
Summary of Contents for Princeton Instruments NIRvana HS
Page 1: ...www princetoninstruments com NIRvana HS System Manual 4411 0169 Issue 1 0 January 8 2020 ...
Page 3: ...Issue 1 0 NIRvana HS System Manual 3 This page is intentionally blank ...
Page 4: ...4 NIRvana HS System Manual Issue 1 0 ...
Page 10: ...10 NIRvana HS System Manual Issue 1 0 ...
Page 26: ...26 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 34: ...34 NIRvana HS System Manual Issue 1 0 ...
Page 52: ...52 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 72: ...72 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 92: ...92 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Page 101: ...This page is intentionally blank ...