Chapter 7
Exposure
51
When vibration may affect results, fan operation can be turned off while making sure
that the coolant is circulating through the camera to maintain the sensor cooling
temperature. If the fan is turned off and there is no coolant circulating through the
camera, the built-in thermo-protection switch may shut the camera down to prevent
thermal damage. In the event that this occurs, wait approximately ten minutes, correct
the situation that caused the shutdown, and re-power the camera.
The deepest operating temperature for a system depends on the sensor size and
packaging. Refer to
Table A-3, Operating Temperatures
, on page 80 for default cooling
temperatures.